The water meter testing benches LS-6B – DN40-150 is designed strictly according to the following standard:
GB-T778.1-1996:Specification
GB-T778.2-1996:Install requirement
GB-T778.3-1996:Test methods and test equipment
JJG162-2007:Verification Regulation
The
bench is available to use the volume comparison method to test water
meters DN40mm, DN50mm, DN80mm, DN100mm and DN150mm.The equipment can
supply whole bench characters analysis and errors inspection to a single
water meter. To ensure the equipment smoothly operation, Mechanical
method only and mechanical combined with cameras are selective ways, the
later gives more convenient. Products are offered to the manufacturer
of the water meter, repair factory, measure department,
large-and-middles-scale industrial and mining enterprises, and railway
water and electricity ones. The product is well known by the features of
High precision, easily operation and simple maintenance.
1. Device accuracy: 0.2
2. Device calibration method: the volume comparison.
3. Standard
graduate cylinder: gourd-style double-necking with 2 gauges, the lower
half of one gauge have a volume of 0.1m3 and the whole is 0.5m3; another
gauge has the lower half 1m3 and the whole 3m3. The necking part of the
standard measure of the corresponding volume of water per liter higher
than the national standard height.
4.
Instantaneous flow meter: measured precision ≤ level 2.5, is divided
into four glass rotor flow meter, one for the shunt orifice flow meter:
4.1 LZB-25,70-700L/h,DN15 ball valve control.
4.2 LZB-50,450-8000L/h,DN25 ball valve control.
4.3 LZB-80,7-35 m3/h,DN50 ball valve control.
4.4 LZB-25,40-180m3/h,DN150
Worm manually control valve , a ball valve DN25 for the flow
fine-tuning purposes next to shunt orifice flow meter..
5. Clamp meter devices: pneumatic, telescopic distance: about 240mm.
6. Commutator: piston.
Trip distance: 180mm.
air source pressure: 0.35~0.5MPa.
When
pressure of gas source≥ 0.4MPa, the time tolerance of the Commutator
movement is less than 0.1% as one shortest flow time go.
7. Water pressure: 0.25~0.4MPa.
Water
pressure should be no discernible pulse phenomena, the slow changes in
water pressure caused by fluctuations in the instantaneous flow
coefficient is not more than 1%.
8. Power: 380V, 50HZ three-phase four-wire.
9. Calibration dimensions of the test bench are nearly: L × b × h is about 7400 × 2800 × 3400mm.
10. Total weight of bench: 3t.